_PCS4RATEUNIT?
Returns the pressure rate of change setting (units
per second or units per minute) if function is
available.
Response: ‘sp digit, name cr lf
_PCS4READING?
Returns a reading from the PCS400 in output
format 1.
Response: ‘sp value cr lf
_PCS4SOURCEP?
Returns the value of the last source pressure test,
or zero if test has not been performed.
Response: ‘sp value cr lf
_PCS4SPAN?
Returns the stored multiplication factor used to
proportionally correct span offset overrange of the
unit.
Response: ‘sp value cr lf
_PCS4STABLEDELAY?
Returns the number of readings that must be
within the stable window before a stable pressure
is indicated. Maximum is 999, default is 67.
Response: ‘sp digits cr lf
_PCS4STABLEWINDOW?
Returns the pressure tolerance allowed for a sta-
ble pressure indication. Default is 0.004% FS.
Response: ‘sp value cr lf
_PCS4STAT?
Returns mode and stability flag.
Response: ‘mode, stable cr lf
_PCS4UNIT?
Returns the number of the current units of pres-
sure measurement, the full name with tempera-
ture, and the configuration of the sensor
(absolute, gauge, differential).
Response: ‘sp digits, name, type cr lf
_PCS4XDUCER?
Returns the number of the currently active
transducer.
Response: ‘sp digits cr lf
_PCS4XDUCERID?
Returns the internal transducer number, serial
number, minimum and maximum sensor range.
Response: ‘sp Mensor,PBT,xducer,sn,min,max
cr lf
_PCS4ZERO?
Returns the stored zero offset of the active
transducer in the current pressure units. (Limited
to values between +16 to –16 psi, or equivalent.)
Response: ‘sp value cr lf
REMOTE OPERATION PCS 400