EasyManua.ls Logo

mensor PCS 400 - Page 47

mensor PCS 400
118 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
_PCS4•RATEUNIT?
Returns the pressure rate of change setting (units
per second or units per minute) if function is
available.
Response: ‘sp digit, name cr lf ’
_PCS4•READING?
Returns a reading from the PCS400 in output
format 1.
Response: ‘sp value cr lf ’
_PCS4•SOURCEP?
Returns the value of the last source pressure test,
or zero if test has not been performed.
Response: ‘sp value cr lf ’
_PCS4•SPAN?
Returns the stored multiplication factor used to
proportionally correct span offset overrange of the
unit.
Response: ‘sp value cr lf ’
_PCS4•STABLEDELAY?
Returns the number of readings that must be
within the stable window before a stable pressure
is indicated. Maximum is 999, default is 67.
Response: ‘sp digits cr lf ’
_PCS4•STABLEWINDOW?
Returns the pressure tolerance allowed for a sta-
ble pressure indication. Default is 0.004% FS.
Response: ‘sp value cr lf ’
_PCS4•STAT?
Returns mode and stability flag.
Response: ‘mode, stable cr lf ’
_PCS4•UNIT?
Returns the number of the current units of pres-
sure measurement, the full name with tempera-
ture, and the configuration of the sensor
(absolute, gauge, differential).
Response: ‘sp digits, name, type cr lf ’
_PCS4•XDUCER?
Returns the number of the currently active
transducer.
Response: ‘sp digits cr lf ’
_PCS4•XDUCERID?
Returns the internal transducer number, serial
number, minimum and maximum sensor range.
Response: ‘sp Mensor,PBT,xducer,sn,min,max
cr lf ’
_PCS4•ZERO?
Returns the stored zero offset of the active
transducer in the current pressure units. (Limited
to values between +16 to –16 psi, or equivalent.)
Response: ‘sp value cr lf ’
REMOTE OPERATION PCS 400
4-6 www.mensor.com

Table of Contents