COMMON REMOTE COMMANDS (Table 9.8 - Quick Reference List)
All the commands begin with “_PCS4" and end with the selected termination character:
Commands:
AUTORANGE 0
AUTORANGE 1
CAL A/D <digit> - xducer no.
CAL ATM
CAL DISABLE OFF
CAL DISABLE ON
CAL SPAN value - true span
CAL ZERO value - true zero
CTRLMAX value - max ctrl press
CTRLMIN value - min ctrl press
CTRL value - ctrl pressure
DEFAULT
FILTERSETTING digits - 00 to 99 filter %
FILTERWINDOW value - max window
FUNC CTRL <value<unitno>>
FUNC MEAS <unitno>
FUNC STBY <unitno>
FUNC VENT <unitno>
FUNC F1 - special function 1
FUNC F2 - special function 2
FUNC F3 - special function 3
OUTFORM digit - 1 ‘sp value crlf’
2 ‘sp value, unitno, function crlf’
3 ‘sp value, ratevalue crlf’
4 ‘sp value, minpeak, maxpeak crlf’
5 ‘sp value, auxvalue crlf’
6 ‘sp value, value, stable crlf’
7 ‘sp value, no barometer crlf’ or
‘sp value, value crlf’
PEAKRESET
PEAKUNIT digit - 0 normal
1 max press
2 min press
RATE value - ctrl rate
RATEUNIT digit - 0 normal
1 units/min
2 units/sec
STABLEDELAY digits - no. of readings for stable
STABLEWINDOW value - pressure window for stable
UNIT unitno
XDUCER digits - active xducer number
Tests:
TEST EXHAUSTP? - ‘T value units crlf’
TEST EXTLEAK? press, seconds - ‘T rate crlf’
TEST INTLEAK? press, seconds - ‘T rate crlf’
TEST MEMORY? - ‘T 1 = chksum, 2 = chksum,
3 = chksum, 4 = chksum, 5 = chksum,
6 = chksum crlf’
TEST PROGRAM? - ‘T checksum crlf’
TEST REG? - ‘T stat value,stat value,stat value,
stat value crlf’ value = press change
TEST SOLENOIDS? - ‘Tstat, stat, stat, stat,
stat, stat crlf’
TEST SOURCEP? - ‘T value units crlf’
TEST XDUCER? <digits> - xducer number - ‘T
stat, pcounts, tcounts crlf’
Queries:
AUTORANGE?
CTRL? - ‘sp value crlf’
CTRLMAX? - ‘sp value crlf’
CTRLMIN? - ‘sp value crlf’
EMUL? - ‘0 crlf’ or ‘1 crfl’
ERR? - ‘Ennnn error string crlf’
EXHAUSTP? - ‘sp value, units crlf’
FILTERSETTING? - ‘sp percent crlf’
FILTERWINDOW? - ‘sp window crlf’
ID? - ‘MENSOR, PCS-400, snsnsn, x.xx crlf’
LIST? - ‘sp 0,1,2 crlf’ or ‘sp 0,2 crlf’
OUTFORM? - ‘sp digits crlf’
PEAKUNIT? - ‘sp digit, name crlf’
RANGEMAX? - ‘sp value crlf’
RANGEMIN? - ‘sp value crlf’
RATE? - ‘sp value crlf’
RATEUNIT? - ‘sp digit, name crlf’
READING? - ‘sp value crlf’
SOURCEP? - ‘sp value, units crlf’
SPAN? - ‘sp value crlf’
STABLEDELAY? - ‘sp no_of_readings crlf’
STABLEWINDOW? - ‘sp value crlf’
STAT? - ‘mode, stable crlf’
UNIT? - ‘sp digits, unitname, abs_or_diff crlf’
XDUCER? - ‘sp digits crlf’
XDUCERID? - ‘sp Mensor, PBT, xducer, sn,
min, max crlf’
ZERO? - ‘sp value crlf’
APPENDIX PCS 400