_PCS4•TEST•SOLENOIDS?
Performs a functional test of the internal pressure
control valves.
Response:‘T stat, stat, stat, stat, stat<,stat> cr lf’
Where stat = PASS or FAIL.
If six solenoids are installed, a sixth stat is re-
turned.
_PCS4•TEST•SOURCEP?
Determines the source pressure applied on the
Supply port of the instrument. (Used to determine
the approximate maximum control pressure.)
Response: ‘T value cr lf ’
Where value equals the approximate source pres-
sure up to a maximum of 110% of the primary
transducer range and units identifies the units of
measure.
_PCS4•TEST•XDUCER?<•digits>
Functionally verifies main internal transducer un-
less other transducer is specified by digits. Tests
for nominal values, not for accuracy.
Response: ‘T stat, digits, digits cr lf ’
Where stat is PASS or FAIL (the result of the RAM
test and digits are the pressure counts and digits
are the temperature counts. If no transducer is
specified, the main transducer (0) is tested. If a
non-existent transducer is specified, the test will
be performed, but will fail with unusable counts
returned.
Queries
_PCS4•AUTORANGE?
Returns 0 cr lf if in Range hold, or
1 cr lf if in Autorange.
_PCS4•CTRL?
Returns the current control point in the current
engineering units.
Response: ‘sp value cr lf ’
_PCS4•CTRLMAX?
Returns the current maximum control pressure
that can be commanded.
Response: ‘sp value cr lf ‘
_PCS4•CTRLMIN?
Returns the current minimum control pressure
that can be commanded.
Response: ‘sp value cr lf ’
_PCS4•EMUL?
Returns the current mode of operation.
Response: ‘0 cf lf’ if in native mode, or
‘1 cr lf’ if in emulation mode.
_PCS4•ERR?
Returns the error number and error description
(see ‘Error Codes’). Clears error indication and
GPIB service request.
Response: ‘Ennnn error_string cr lf ’
Where ‘nnnn’ is the error number, and er-
ror_string is the description terminated with a
carriage return and linefeed.
_PCS4•EXHAUSTP?
Returns the value of the last exhaust pressure test
or zero if test has not been performed.
Response: ‘T sp value cr lf ’
_PCS4•FILTERSETTING?
Returns the current percentage of filter.
Response: ‘sp digits cr lf ’
_PCS4•FILTERWINDOW?
Returns the value of the filter window.
Default is 0.025% FS.
Response: ‘sp value cr lf ’
_PCS4•ID?
Returns instrument ID consisting of: MEN-
SOR,PCS-400,(six digit serial number nnnnnn),
(4 character version number n.nn)
_PCS4•LIST?
Returns a list of the sensors installed in instru-
ment.
Response: ‘sp 0,1,2 cr lf’ for a PCS400 with 3
internal sensors,or ‘sp 0,2 cr lf’ for a primary and
a barometric reference.
_PCS4•OUTFORM?
Returns the number of the current output format
(see PCS 400 Command Responses, Table 4.1).
Response: ‘sp digits cr lf ’
_PCS4•PEAKUNIT?
Returns the units of measure of the peak reading.
Response: ‘sp digit, name cr lf ’
_PCS4•RANGEMAX?
Returns the maximum pressure of the currently
active transducer.
Response: ‘sp value cr lf ’
_PCS4•RANGEMIN?
Returns the minimum pressure of the currently
active transducer.
Response: ‘sp value cr lf ’
_PCS4•RATE?
Returns the rate of pressure change setting.
Response: ‘sp value, digit, mode cr lf ’
value = pressure rate in units per second
digit = 0 for fast, 1 for rate
mode = FAST or RATE
PCS 400 REMOTE OPERATION
www.mensor.com 4-5