EasyManua.ls Logo

METER GROUP AquaLab 4TE - Page 32

METER GROUP AquaLab 4TE
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
28
SYSTEM
Calibration. Select to start the calibration process (Section4.1) for water activity or
moisture content (Figure37). Resetting factory defaults is described in Section4.1.4.
Figure37 Calibration option
Temperature. Select to set the AQUALAB to the desired temperature using UP and DOWN.
When finished, select SAVE (Figure38).
The AQUALAB 4TE models may be set between 15 and 50 °C in 0.1 °C intervals. The default
temperature is 25 °C.
Figure38 Temperature option
Temp Eq. Select to set the acceptable temperature difference between the sample block
and the sample before starting a water activity measurement (Figure39).
The range is 0.1 to 4.0 °C. A setting of 4.0 °C begins the measurement when the sample is
<4.0 °C above or below the block temperature. A setting of 0.1 °C begins the measurement
when the sample is <0.1 °C above or below the block temperature.

Related product manuals