MI 3125 InstaltestCOMBO Voltage, frequency, phase sequence
26
5.2 Voltage, frequency and phase sequence
Voltage and frequency measurement is always active in the terminal voltage monitor. In
the special VOLTAGE TRMS menu the measured voltage, frequency and information
about detected three-phase connection can be stored. Phase sequence measurement
conforms to the EN 61557-7 standard.
See chapter 4.1 Function selection for instructions on
key functionality.
Figure 5.4: Voltage in single
phase system
Test parameters for voltage measurement
There are no parameters to set.
Connections for voltage measurement
L3
L2
L1
N
PE
N
/
L
2
N
/
L
2
L
/
L
1
L
/
L
1
P
E
/
L
3
P
E
/
L
3
result 1.2.3 result 2.1.3
Figure 5.5: Connection of universal test cable and optional adapter in three-phase
system
L1
L2
L3
N
PE
R
E
Ro
LPEN
L
/
L
1
N
/
L
2
P
E
/
L
3
N/L2
L
/
L
1
PE/L3
Figure 5.6: Connection of plug commander and universal test cable in single-phase
system

Table of Contents

Other manuals for METREL MI 3125

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the METREL MI 3125 and is the answer not in the manual?

METREL MI 3125 Specifications

General IconGeneral
TypeMultifunction Installation Tester
Insulation ResistanceYes
Continuity TestYes
Loop ImpedanceYes
RCD TestingYes
Voltage MeasurementYes
Frequency MeasurementYes
DisplayLCD
Power Supply6 x 1.5 V AA batteries
Loop Resistance Resolution0.01 Ω
Frequency45 - 65 Hz
Line Voltage Resolution1 V
Insulation Resistance Test Voltage50 V, 100 V, 250 V, 500 V, 1000 V
Insulation Resistance Resolution0.01 MΩ
Continuity Resistance Resolution0.01 Ω
Test Current200 mA
DC VoltageUp to 600 V
Loop Resistance0.0 Ω - 1999 Ω
Line VoltageUp to 550 V
Line Voltage Accuracy±2% ±2 digits
Rated Residual Current10 mA, 30 mA, 100 mA, 300 mA, 500 mA
Insulation Resistance Range0.01 MΩ to 1 TΩ
Continuity Resistance0.0 Ω - 1999 Ω
Dimensions220 x 103 x 55 mm
Operating Temperature-10°C to 50°C
Storage Temperature-20°C to +60°C
Safety StandardsEN 61557
Loop Resistance Accuracy±(5% + 3 digits)