EasyManuals Logo

Metrohm 756 KF User Manual

Metrohm 756 KF
163 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #87 background imageLoading...
Page #87 background image
4.2 Remote control commands
756/831 KF Coulometer, Instructions of Use
79
Config.Monitoring.Reagent.ReagCap 1...1000...9999, OFF
Config.Monitoring.Reagent.RCounter 0...9999
Config.Monitoring.Reagent.ClearCount $G
Config.Monitoring.Reagent.Drift 0...99, OFF
Monitoring of reagent live time.
.Determ: Number of determinations.
.DCounter: Counter of determinations already carried out.
.MaxTime: Maximum live time of reagent in days.
.TCounter: Time already elapsed since last reagent change.
.ReagCap: Reagent capacity in mg water.
.RCounter: Counter of reagent capacity.
.ClearCount: Clears all above counters.
.Drift: Stable drift in ug/min.
4.2.2.32. Config.Monitoring.Change $G, $S
Config.Monitoring.Change.Status auto, man., OFF
Config.Monitoring.Change.WaitTime 0...999 999
Config.Monitoring.Change.AspVol 0...100...9999
Config.Monitoring.Change.SolventVol 0...100...9999
Config.Monitoring.Change.Rinse 0...9999
Config.Monitoring.Change.NoRinse 1...9
Changing of reagent. With a connected Dosino, the reagent is changed with
&Config.Monitoring.Change $G. The nod &Config.Monitoring.change.Status
has to be OFF.
Parameters for automatic reagent change:
.WaitTime: Waiting time in s after switching off the stirrer.
.AspVol: Volume in ml of used reagent to be aspirated.
.SolventVol: Volume in ml of new reagent to be added.
.Rinse: Volume in ml of rinsing reagent.
.NoRinse: Number of rinsing cycles.
4.2.2.33. Config.Monitoring.Validation.Status ON, OFF
Config.Monitoring.Validation.Interval 1...365...9999
Config.Monitoring.Validation.Counter 0...9999
Config.Monitoring.Validation.ClearCount $G
Monitoring of validation.
.Interval: Time interval in days for validation.
.Counter: Time counter in days since last validation.
.ClearCount: Clears the above counter.
4.2.2.34. Config.Monitoring.Service.Status ON, OFF
Config.Monitoring.Service.Date XXXX-XX-XX
Monitoring of service interval.
4.2.2.35. Config.Monitoring.DiagRep ON, OFF
Printing of system test report after each switching on of the Coulometer.
4.2.2.36. Config.PeriphUnit.CharSet1 Epson, Seiko, Citizen, HP, IBM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 756 KF and is the answer not in the manual?

Metrohm 756 KF Specifications

General IconGeneral
BrandMetrohm
Model756 KF
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals