EasyManuals Logo

Metrohm 756 KF User Manual

Metrohm 756 KF
163 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #89 background imageLoading...
Page #89 background image
4.2 Remote control commands
756/831 KF Coulometer, Instructions of Use
81
Current sample number.
Set to 0 on power on and initialization. After 9999, counting starts again at 0.
4.2.2.44. Config.Aux.OpLevel standard, expert
Operator level for manual operation.
4.2.2.45. Config.Aux.StartDelay 0...999 999
Start delay time in s. During this time, the data of the preceding determination
are retained.
4.2.2.46. Config.Aux.ResDisplay bold, standard
Character set for the result display at the end of the determination.
4.2.2.47. Config.Aux.DevName up to 8 ASCII characters
Name of the instrument for connections with several units. It is advisable to
use only the letters A...Z (ASCII No. 65...90), a...z (ASCII No. 97...122) and
the numbers 0...9 (ASCII No. 48...57) when the function Setup.AutoInfo
(4.2.2.97) is used at the same time.
If a name has been entered, it will be printed out in the result report (full,
short).
4.2.2.48. Config.Aux.Beep 1...3, OFF
Number of beep sounds.
4.2.2.49. Config.Aux.DisplayMeas ON, OFF
Display of potentials during conditioning and titration.
4.2.2.50. Config.Aux.Prog read only
Output of the program version.
The Coulometer sends "5.756.0010" on requests with $Q.
4.2.2.51. Config.RSSet1 $G
Config.RSSet1.Baud 300, 600, 1200, 2400, 4800, 9600
Config.RSSet1.DataBit 7, 8
Config.RSSet1.StopBit 1, 2
Config.RSSet1.Parity even, odd, none
Config.RSSet1.Handsh HWs, SWchar, SWline, none
$G sets all RS settings. The changes are performed only if the instrument is
inactive. After the setting of the interface parameters, wait at least 2 s to allow
the components to equilibrate.
Settings of the values for the data transmission via the RS interface: baud
rate, data bit, stop bit, parity and type of handshake, see also page 97 ff.
The setting of the values must be initiated with $G immediately after entry of
the values.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 756 KF and is the answer not in the manual?

Metrohm 756 KF Specifications

General IconGeneral
BrandMetrohm
Model756 KF
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals