EasyManua.ls Logo

Metrohm 774 - Page 47

Metrohm 774
168 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.3 KF Moisture determination
774 Oven Sample Processor, Instructions for use
41
water
[
µ
g]
t [min]
vial humidity
in
µ
g
system drift
in
µ
g/min
determination interval
total blank value
drift
[µg H
2O/min]
t [min]
min. extraction time
blank
sample
min. extraction time
determination interval
The blank value is made up of the moisture
which adheres to the sample vial * and a sys-
tem drift which is latently present and whose
cause can be found in a minimal residual mois-
ture in the carrier gas and possibly leaks in the
tubing system.
*
The use of pre-dried sample vials has no significant advan-
tages. On the contrary, this causes the reproducibility of
the results to be adversely affected.
The accompanying diagram shows the theoreti-
cal course of a sample determination with the
774 Oven Sample Processor. It is clear that, in
order to determine the amount of water correctly
(dark gray area), the blank value must be sub-
tracted from the total amount of moisture deter-
mined.
Sample determination
When the reaction medium has been conditioned fully the gas flow is in-
terrupted, the sample moved to the oven and introduced into the oven
block with the help of the penetration needle.
The gas flow is switched on again and the sample determination started
at the Coulometer or KF Titrator. Definition of an extraction time ensures
that a minimum determination time and therefore heating period will be
observed.
The driven-off moisture is now transported to the reaction vessel by the
stream of gas. The transfer tube heating ensures that no moisture can
condense in the tubing.
When the determination is finished the moisture content of the sample is
calculated taking the previously determined blank value into considera-
tion.
The gas flow is switched off again and the conditioning vessel again
move to the oven.
When the needle has penetrated the conditioning vessel septum the gas
flow is switched on again and the reaction medium is conditioned.
When conditioning has been carried out the next sample can be deter-
mined, see above.
Care must be taken that the septa are still airtight after penetration by the
hollow needle and that no moisture can escape. The septum of the condi-
tioning vessel is particularly at risk as it is penetrated many times during the
sample series.

Table of Contents

Related product manuals