EasyManua.ls Logo

Metrohm 774 - Initialise Data Memory

Metrohm 774
168 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6.5 Initialise data memory
774 Oven Sample Processor, Instructions for use
149
Magnets
fitted
In this
example, the code is:
000110
Magnet fixture viewed from below
Raise the rack and place it to one side.
Prepare the unit for diagnosis (see Chap. 6.4.2).
If necessary, press<Ð> several times until
diagnosis
>power on reset
<ENTER>
The unit runs through the power-up routine (initialisation of lift and rack positions). Initialisation is
important since it is essential that the turntable (rack holder) be in basic position (beaker 1 at
tower 1) for the following “rack code test“ diagnosis step.
Prepare the unit for diagnosis (see Chap. 6.4.2).
If necessary, press <Ð> several times until
diagnosis
>rack code test
<ENTER>
The test reads in the coding constantly and displays it. A six-digit bit pattern (code ??????) is
scheduled for this for the display. The first digit for magnet No. 1, the second digit for magnet
No. 2 etc. If a magnet is detected, the corresponding digit is assigned "1". Otherwise, it is as-
signed "0".
Consecutively fit all racks to be used and compare the related note
which you made beforehand with the information on the display.
>rack code test
code 000000
Example of:
No rack fitted
>rack code test
code 000001
Example of:
Rack with coding in accordance with above
example.
You can quit the test by pressing key <QUIT> or <STOP>.
diagnosis
>power on reset
You can quit diagnosis from the main menu with <QUIT> or <STOP>.
6.5 Initialise data memory
This diagnosis step allows you to assign default values to device parame-
ters using the keypad and thus set the unit to its original state. This is im-
portant owing to the following two aspects:

Table of Contents

Related product manuals