5.10 The remote control tree
774 Oven Sample Processor, Instructions for use
109
:
├ .WetPart Dosing unit definitions - 5.11.2.9
│ ├ .WetPartNo Dosing unit no. 1...12
│ ∗
│ ├ .1 Port number 1 - 5.11.2.10
│ │ ├ .MaxRate Max. dosing rate 0.01...160 ml/min
│ │ ├ .Length Tubing length 0...1000...30000 mm
│ │ └ .Diameter Tubing diameter 0...2...20 mm
│ │
│ ├ .2 Port number 2 - 5.11.2.10
│ │ ├ .MaxRate Max. dosing rate 0.01…160 ml/min
│ │ ├ .Length Tubing length 0...250...30000 mm
│ │ └ .Diameter Tubing diameter 0.1…2…20 mm
│ │
│ ├ .3 Port number 3 - 5.11.2.10
│ │ ├ .MaxRate Max. dosing rate 0.01...160 ml/min
│ │ ├ .Length Tubing length 0...1000...30000 mm
│ │ └ .Diameter Tubing diameter 0.1...2...20 mm
│ │
│ └
.4 Port number 4 5.11.2.10
│ ├ .MaxRate Max. dosing rate 0.01...160 ml/min
│ ├ .Length Tubing length 0...1000...30000 mm
│ └ .Diameter Tubing diameter 0.1...2...20 mm
│
└ .Rsset RS232 Settings $G 5.11.2.11
├ .Baud Baud rate 300, 600, 1200, 5.11.2.12
│ 2400, 4800, 9600
├ .DataBit Number of data bits 7, 8
├ .StopBit Number of stop bits 1, 2
├ .Parity Parity even, odd, none
├ .Handsh Handshake Hws, HWfull,
│ SWchar, SWline,
│ none
└ .CharSet Character set IBM, HP, Epson, 5.11.2.13
Seiko, Citizen
5.10.4 &Info
Object Description Input range Reference
& Root
:
├ Info Current data -
: ├ .Report Report definition $G 5.11.3.1
│ └ .Select Report type short, full, config, param,
│ usermeth, all
│
└ .DetermData Determination data - 5.11.3.2
├ .Mplist Meas. point list of temperature -
: ├ .1 Meas. point 1 -
: └ .Y Temperature in °C read only
:
└.400 Meas. point 400