5.10 The remote control tree
774 Oven Sample Processor, Instructions for use
113
: :
.Heater Heater messages -
.T Changes of target temperature on, off
.S When heater is switched off on, off
.E When an error occurs on, off
.PowerOn RESET (power on) $G 5.11.4.10
.Initialize Set default values $G 5.11.4.11
.Select Selection of branch param, config,
assembly, setup, all
.RamInit Initialization of working mem. $G 5.11.4.12
.InstrNo Instrument number - 5.11.4.13
.Value Description 8 ASCII characters
(not available in manual operation)
5.10.6 &UserMeth
Object Description Input range Reference
& Root
:
UserMeth User defined methods
: .FreeMemory Memory available read only 5.11.5.1
.Recall Load method $G 5.11.5.2
.Name Method name 8 ASCII characters
.Store Save method $G
.Name Method name 8 ASCII characters
.Delete Delete method $G
.Name Method name 8 ASCII characters
.DelAll Delete all methods $G
.List List of methods - 5.11.5.3
.1 Method 1
.Name Method name read only
.Bytes Method size in bytes read only
5.10.7 &Assembly
Object Description Input range Reference
& Root
:
Assembly Assembly control
: .Sample Beaker position $G 5.11.6.1
.Func Selection of function =, +, –
.Value Value of function 0...1...999
.Move Turning the rack $G, $S 5.11.6.2
.Target Lift selection 1, 2
.Position Beaker or position sample, spec.1...8, 1...999
: