5 Detailed description
774 Oven Sample Processor, Instructions for use
128
5.11.6 &Assembly
5.11.6.1 Assembly.Sample $G
Assembly.Sample.Func =, +, –
Assembly.Sample.Value 0...1...999
Defines the (first) sample beaker (rack position) to be processed. Modifica-
tion of the sample variable.
.Func Function
.Value Value (absolute or relative)
&Assembly.Sample;$G triggers this function.
5.11.6.2 Assembly.Move $G, $S
Assembly.Move.Target 1, 2
Assembly.Move.Position sample, spec.1…8, 0...999
Positioning a beaker at the specified tower.
.Target Target or tower
.Position Rack position or identification of a specific vial
&Assembly.Move;$G triggers this function.
5.11.6.3 Assembly.Lift $G, $S
Assembly.Lift.Station 1, 2,
Assembly.Lift.Way rest, work, rinse, shift, special, 0...325 mm
Move lift.
.Station Selection of lift station (
= both lifts)
.Way absolute lift position
&Assembly.Lift;$G triggers this function.
5.11.6.4 Assembly.Heater $G
Assembly.Heater.Temp Init, off, 50…250
Assembly.Heater.Time blank, 1...999
Temperature control.
.Temp Target temperature in °C
.Time Heating interval in minutes
&Assembly.Heater;$G triggers this function.
5.11.6.5 Assembly.Flow $G
Assembly.Flow.Address Pump, Valve
Assembly.Flow.State on, off
Pump or valve control.
.Address Selection of pump or valve
.Value State
&Assembly.Flow;$G triggers this function.
5.11.6.6 Assembly.Dos $G, $S
Assembly.Dos.Address 1...12
Assembly.Dos.Value ±0.01…999.999 ml, fill, release,
prepar., empty , eject, adjust, level
Control of dosing units.
.Address Selection of dosing unit
.Value Volume or function
&Assembly.Dos;$G triggers this function.