5.5 Printing reports
774 Oven Sample Processor, Instructions for use
85
5.5 Printing reports
You can have reports printed out directly by the 774 Oven Sample Proces-
sor for documentation purposes. However, this does necessitate connect-
ing a printer with serial port to the RS232 socket. This means that no direct
RS232 connection to a Metrohm Coulometer 756 is possible. Please refer
to Page 20f for details of how to connect and configure a printer.
The available reports 774 Oven Sample Processor are as follows:
• Full result report
— Comprehensive listing of oven temperatures, gas flow rates and error
list
• Short result report
— Short listing of the oven temperatures and gas flow rates
• Configuration report
— Listing of all configuration entries
• Parameter report
— Listing of all parameters of a method
• User method report
— Listing of the names of all stored methods
Examples of the reports:
Short result report
'sr
774 Oven Sample Proc. 0130/03 5.774.0010
run number 1
method 756Pump
oven temp. 120.0 °C
lowest temp. 120.1 °C
highest temp. 124.8 °C
mean gas flow 59.0 mL/min
============
Å Report identification ('sr = short report)
Å Report title with device number and program version
Å Sample number
Å Method name
Å Mean oven temperature
Å Minimum oven temperature
Å Maximum oven temperature
Å Mean gas flow rate
The values listed each refer to the part-sequence currently being processed or processed last
(start sequence, sample sequence or final sequence) of the method.
Full result report
'fr
774 Oven Sample Proc. 0130/03 5.774.0010
run number 1
method 756Pump
oven temp. 120.0 °C
lowest temp. 119.7 °C
highest temp. 122.7 °C
mean gas flow 59.0 mL/min
lowest flow 56.7 mL/min
highest flow 61.7 mL/min
smpl heating time 226 s
>error list
11 s * gas flow too low
>table time temperature
Å Report identification ('fr = full report)
Å Report title with device number and program version
Å Sample number
Å Method name
Å Mean oven temperature
Å Minimum oven temperature
Å Maximum oven temperature
Å Mean gas flow rate
Å Minimum gas flow rate
Å Maximum gas flow rate
Å Dwell time of sample in oven
Å List of errors which have occurred