5.5 Printing reports
774 Oven Sample Processor, Instructions for use
85
5.5 Printing reports
You can have reports printed out directly by the 774 Oven Sample Proces-
sor for documentation purposes. However, this does necessitate connect-
ing a printer with serial port to the RS232 socket. This means that no direct
RS232 connection to a Metrohm Coulometer 756 is possible. Please refer
to Page 20f for details of how to connect and configure a printer.
The available reports 774 Oven Sample Processor are as follows:
Full result report
Comprehensive listing of oven temperatures, gas flow rates and error
list
Short result report
Short listing of the oven temperatures and gas flow rates
Configuration report
Listing of all configuration entries
Parameter report
Listing of all parameters of a method
User method report
Listing of the names of all stored methods
Examples of the reports:
Short result report
'sr
774 Oven Sample Proc. 0130/03 5.774.0010
run number 1
method 756Pump
oven temp. 120.0 °C
lowest temp. 120.1 °C
highest temp. 124.8 °C
mean gas flow 59.0 mL/min
============
Å Report identification ('sr = short report)
Å Report title with device number and program version
Å Sample number
Å Method name
Å Mean oven temperature
Å Minimum oven temperature
Å Maximum oven temperature
Å Mean gas flow rate
The values listed each refer to the part-sequence currently being processed or processed last
(start sequence, sample sequence or final sequence) of the method.
Full result report
'fr
774 Oven Sample Proc. 0130/03 5.774.0010
run number 1
method 756Pump
oven temp. 120.0 °C
lowest temp. 119.7 °C
highest temp. 122.7 °C
mean gas flow 59.0 mL/min
lowest flow 56.7 mL/min
highest flow 61.7 mL/min
smpl heating time 226 s
>error list
11 s * gas flow too low
>table time temperature
Å Report identification ('fr = full report)
Å Report title with device number and program version
Å Sample number
Å Method name
Å Mean oven temperature
Å Minimum oven temperature
Å Maximum oven temperature
Å Mean gas flow rate
Å Minimum gas flow rate
Å Maximum gas flow rate
Å Dwell time of sample in oven
Å List of errors which have occurred

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 774 and is the answer not in the manual?

Metrohm 774 Specifications

General IconGeneral
BrandMetrohm
Model774
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals