5 Detailed description
774 Oven Sample Processor, Instructions for use
86
0 s 120.9 °C
10 s 121.9 °C
20 s 122.7 °C
30 s 121.9 °C
40 s 120.7 °C
50 s 120.5 °C
... ...
350 s 121.1 °C
360 s 121.1 °C
370 s 120.7 °C
============
Å Oven temperatures, intervals in accordance with
'recording interval'
The values listed each refer to the part-sequence currently being processed or processed last
(start sequence, sample sequence or final sequence) of the method
Configuration report
'co
774 Oven Sample Proc. 0130/03 5.774.0010
configuration
>auxiliaries
dialog: english
display contrast 3
beeper: ON
device label M774-1
program 5.774.0010
max. lift way 90 mm
beaker sensor: ON
>oven settings
initial temp.: OFF °C
max.temperature 275 °C
temp.correction 0 °C
>rack definitions
number code type
1 000001 M36-0
work position 88 mm
rinse position 48 mm
shift position 0 mm
special position 0 mm
position special beaker 1...8
36 0 0 0 0 0 0 0
>dosing units
default values
tube max.rate length diameter
1 160 mL/min 1000 mm 2.0 mm
2 160 mL/min 250 mm 2.0 mm
3 160 mL/min 1000 mm 2.0 mm
4 160 mL/min 30 mm 2.0 mm
changed values
no tube max.rate length diameter
>RS232 settings
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: HWs
character set: IBM
RS control: ON
------------
Å Report identification ('co = configuration report)
Å Report title with device number and program version
Å Freely selectable ID designation for the device
Å Release number of the device software (cannot be
changed)
Å Default settings of the standard rack
Å Reserved special positions (conditioning vessel at 36)
Å Standard definitions for a connected Dosino