5 Detailed description
774 Oven Sample Processor, Instructions for use
122
5.11.3 &Info …
5.11.3.1 Info.Report $G
Info.Report.Select full, short, config, param, usermeth, all
$G sends the selected report via RS interface.
full: Full result report (identifier 'fr)
short: Short result report (identifier 'sr)
config: Configuration report (identifier 'co)
param: Parameter or method report (identifier 'pa)
user meth: Listing of method storage (identifier 'um)
all: Full report
Reports, that are sent from the changer are marked with a space (ASCII 32)
and the specific report identifier (see above).
5.11.3.2 Info.DetermData.Mplist.1.Y read only
up to .400.Y
Indexed list of the measuring points of the oven temperature. The time in-
terval between the individual measuring points is defined by node
&Mode.Report.TDelta.
5.11.3.3 Info.DetermData.Errorlist.1.Time read only
Info.DetermData.Errorlist.1.ErrNo read only
up to .10.Time bzw. .10.ErrNo
Up to 10 error messages per sequence are stored in the error list. Besides
the instant of occurrence (from the start of the sequence, in seconds), the
error number (see RS error list Page 104) is also stored.
5.11.3.4 Info.Results.TempSet read only
Info.Results.LowTemp read only
Info.Results.HighTemp read only
Info.Results.Gasflow read only
Info.Results.Lowflow read only
Info.Results.Highflow read only
Info.Results.SmplHeatTime read only
The results each refer to the current sequence.
5.11.3.5 Info.ActualInfo.Meas.CyclNo read only
Info.ActualInfo.Meas.OvenTemp read only
Info.ActualInfo.Meas.Gasflow read only
$Q sends the current, measured values.
5.11.3.6 Info.ActualInfo.Lift.1.Exist read only
Info.ActualInfo.Lift.1.MaxHeight read only
Info.ActualInfo.Lift.1.ActHeight read only
Info.ActualInfo.Lift.1.Beaker read only
Info.ActualInfo.Lift.2.Exist read only
Info.ActualInfo.Lift.2.MaxHeight read only
Info.ActualInfo.Lift.2.ActHeight read only
Info.ActualInfo.Lift.2.Beaker read only
Current data of lift 1 and 2.
Exist: Tower exists (yes/no)
MaxHeight: Preset max. height