5.5 Printing reports
774 Oven Sample Processor, Instructions for use
87
Parameter report / method
'pa
774 Oven Sample Proc. 0130/03 5.774.0010
parameters
method 756Pump
number of samples: rack
>start sequence
1 CTL:Rm: INIT
2 MOVE 1 : spec.1
3 CTL:Rm: START device1
4 HEATER:INIT°C min
5 LIFT: 1 : rinse mm
6 FLOW: pump : ON
7 WAIT 300 s
8 SCN:Rm : Cond ok
>sample sequence
1 MOVE 1 : spec.1
2 LIFT: 1 : rinse mm
3 FLOW: pump : ON
4 WAIT 30 s
5 SCN:Rm : Cond ok
6 WAIT 60 s
7 SCN:Rm : Cond ok
8 FLOW: pump : OFF
9 SCN:Rm : no error
10 CTL:Rm: START device1
11 MOVE 1 : sample
12 LIFT: 1 : work mm
13 FLOW: pump : ON
14 SCN:Rm : End1
15 FLOW: pump : OFF
16 WAIT 5 s
>final sequence
1 RACK
>report
report: none
rec. interval 10 s
>changer settings
rack number 0
lift rate 1 12 mm/s
lift rate 2 12 mm/s
shift rate 20
shift direction: auto.
on beaker error: MOVE
>timeout settings
HEATER timeout: 20 min
on HEATER timeout: STOP
SCAN timeout: 20 min
on SCAN timeout: error
>gas flow
unit gas flow: mL/min
min. flow 50 mL/min
max. flow 100 mL/min
gas type: air
>dosing unit def.
>manual stop
CTL Rmt: **************
CTL RS232:
FLOW: OFF
HEATER: INIT °C
------------
Å Report identification ('pa = parameter report)
Å Report title with device number and program version
Å Method name
Å Scope of sample series
Å Command sequence run once prior to sample series
Å Command sequence for each individual sample
Å Command sequence run after the sample series
Å Rack reset
Å Type of report
Å Changer settings
Å Definition of behaviour in the case of exceptional states
Å Parameters for the carrier gas stream
Å Behaviour in the event of manual method abort
See the Tutorial for further method listings.