5 Detailed description
774 Oven Sample Processor, Instructions for use
110
:
.Errorlist List of error messages - 5.11.3.3
.1 Error message 1 -
: .Time Time of occurence in s read only
.ErrNo Error number read only, see error messages
:
: .10
.Results Resultats - 5.11.3.4
: .TempSet Temperatur set read only
.LowTemp Lowest temperature read only
.HighTemp Highest temperature read only
.Gasflow Mean of gas flow read only
.Lowflow Lowest gas flow read only
.Highflow Highest gas flow read only
.SmplHeatTime Heating interval of sample in min read only
:
.ActualInfo Current data - 5.11.3.5
.Meas Meas. value temperature -
.CyclNo Cycle number read only
.OvenTemp Oven temperature read only
.Gasflow Gas flow rate read only
.Lift Lift station - 5.11.3.6
.1 Lift 1 -
.Exist Availability read only
.MaxHeight Max. stroke path read only
.ActHeight Current lift position read only
.Beaker Presence of beaker read only
.2 Lift 2 -
.Exist Availability read only
.MaxHeight Max. stroke path read only
.ActHeight Current lift position read only
.Beaker Presence of beaker read only
.Rack Sample rack - 5.11.3.7
.Code Rack ID code read only
.Type Rack type read only
.WorkHeight Work position read only
.RinseHeight Rinse position read only
.ShiftHeight Shift position read only
.SpecialHeight Special position read only
.ActPos Current beaker pos. tower 1 read only
.Act2Pos Current beaker pos. tower 2 read only
.Valve Valve - 5.11.3.8
.State State read only
.Pump Pump - 5.11.3.9
.State State read only
.OutHeater Outlet heater - 5.11.3.10
.State State read only
.Fan Fan - 5.11.3.11
.State State read only
: