• 31 •
Setting the LOW VOLTAGE threshold
1. Press the ARROW (or use the numeric keypad) to highlight LOW VOLTAGE, and press
ENTER.
2. Press the
ARROW to scroll to the number of volts.
3. Press
ENTER to accept.
Table 6: Low voltage threshold ranges and defaults
Range Default Range Default
1.000 to 2.750 V for a 2 V jar 2.1 V 5.000 to 13.800 V for a 10 V jar 10.5 V
2.000 to 5.500 V for a 4 V jar 4.2 V
6.000 to 16.500 V for a 12 V jar
12.6 V
3.000 to 8.250 V for a 6 V jar 6.3 V
7.000 to 19.250 V for a 14 V jar
14.7 V
4.000 to 11.000 V for an 8 V jar 8.4 V
8.000 to 22.000 V for a 16 V jar
16.8 V
Selecting a test mode (JARS ONLY)
The default is JARS ONLY.
1. Press the
ARROW keys to highlight JARS ONLY, JARS & 1 STRAP, or JARS & 2 STRAPS in
the STRING INFO list, and press
ENTER.
2. Press the
ARROW keys (or use the numeric keypad) to check the box next to JARS ONLY,
JARS & 1 STRAP, or JARS & 2 STRAPS.
3. Press
ENTER to accept.
Selecting the number of POSTS PER JAR
The default is 2.
1. Press the
ARROW keys to highlight POSTS PER JAR, and press ENTER.
2. Press the
ARROW keys (or use the numeric keypad) to scroll to the number of posts.
3. Press
ENTER to accept.
Chapter 5: Selecting test parametersCelltron ULTRA