33
Selecting what to measure (VOLT & CONDUCTANCE)
The default is VOLT & CONDUCTANCE.
1. Press the
ARROW keys to highlight VOLT & CONDUCTANCE, and press ENTER.
2. Press the
ARROW keys (or use the numeric keypad) to check the box next to your choice:
VOLTAGE ONLY, VOLT & CONDUCTANCE, VOLTS THEN CONDUCT(ANCE).
NOTE: Select VOLTS THEN CONDUCT(ANCE) if you are measuring conductance and do
not want conductance testing to affect your voltage measurements. After you measure the
voltage of the jars (or jars and straps) in a string, disconnect, then reconnect to the rst jar in
the string to complete the test by measuring conductance.
3. Press
ENTER to accept.
NOTE: Once you start the conductance portion of the VOLTS THEN CONDUCTANCE test,
you will not be able retest voltage. Be sure to review your voltage reading before you start
your conductance portion of the test.
Selecting the string TEMPERATURE
The default is 25 °C (77 °F). Conductance measurements vary by jar temperature. To compensate,
the analyzer uses the TEMPERATURE value to adjust the percentage of the reference value you set
in the UTILITIES SETUP as the WARN/FAIL threshold. The percent is compensated to 25 °C (77 °F).
Compensation is adjusted at 0.7% per degree Celsius between 0 °C and 35 °C. (The kit includes an
infrared temperature sensor and instructions.)
1. Press the
ARROW keys to highlight TEMPERATURE, and press ENTER.
2. Press the
ARROW keys (or use the numeric keypad) to select a value.
3. Press
ENTER to accept.
Saving the BATTERY SETUP
To save your settings for later use, press the ARROW keys to highlight SAVE AND CONTINUE, and
press ENTER. You are returned to the NEW test menu. To use a saved BATTERY SETUP to test a
string, see Selecting the TEST option in Chapter 6: Testing.
To save your settings for immediate testing, connect the clamps or probes and press the
ARROW to
highlight SAVE AND TEST. Press
ENTER to start the test. (See Chapter 6: Testing.)
Chapter 5: Selecting test parametersCelltron ULTRA

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Midtronics CTU-6000 and is the answer not in the manual?

Midtronics CTU-6000 Specifications

General IconGeneral
BrandMidtronics
ModelCTU-6000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals