Replacing the Quantum Cartridge, Continued
Rinsing
• The Quantum Cartridge, when newly installed, needs to be rinsed.
• This ensures optimal water quality.
• Make sure that the reservoir is at least 80% full of water.
Step Action Diagram
1
• Locate the clear tubing and
the barbed fitting from the
Milli-Q System Accessories
Bag.
• Screw the barbed fitting onto
the Q-POD Unit.
NOTE:
Do not use any white tape on
the threads of the barbed fitting.
An O-ring is located inside the
Q-POD Unit.
• Push one end of the clear
tubing onto the end of the
barbed fitting.
• Place the other end of the
clear tubing into a sink.
2 The Milli-Q System must be in
READY Mode.
3
• Push the plunger down on the
Q-POD Unit.
• In a few minutes, water
should dispense from the
Q-POD Unit.
Continued on next page