User Configuration Operation
3 - 34 0070-10-0692-02 Accutorr Plus™ Operating Instructions
5b Time Out, LEDs and LCD
Characters when unit is
powered from the
internal battery.
Set how long the numeric
information is displayed,
when no keys have been
pressed, in the LEDs and
LCD before they are
blanked to conserve energy.
The choices are: 5, 15, 20
or 30 minutes. NOTE: The
information is not erased.
5 minutes
5c Time Out, Light in the
LCD when the unit is
powered from AC
mains.
Set how long the light will
stay on, when no keys are
pressed, in the LCD. The
choices are: 3, 5, 10 or 15
minutes.
3 minutes
5d Time Out, Light in the
LCD when the unit is
powered from the
internal battery.
Set how long the light will
stay on, when no keys are
pressed, in the LCD. The
choices are: 3, 5, 10 or 15
minutes.
3 minutes
6a Adult Initial Inflation
Pressure
Set the initial cuff inflation
pressure for an adult size
patient. The choices are:
100 to 260 mmHg at 5
mmHg increments.
180 mmHg
6b Pediatric Initial Inflation
Pressure
Set the initial cuff inflation
pressure for a pediatric size
patient. The choices are: 60
to 180 mmHg at 5 mmHg
increments.
140 mmHg
6c Neonate Initial Inflation
Pressure
Set the initial cuff inflation
pressure for a neonate size
patient. The choices are: 40
to 120 mmHg at 5 mmHg
increments.
100 mmHg
7 Adaptive Inflation Choices are ON or OFF. If
User Configuration #3 is set
to display graphics, the
choices are -l- or -O-.
ON
8 Interval Setting Set the NIBP Interval Time.
The choices are: OFF (or —
—), Cont. (Continuous), 1,
2.5, 5, 10, 15, 20, 30, 60,
120 and 240 minutes.
OFF (or ——)
9a Adult Alarm Limits Set the default alarm limit
values for an Adult size
patient. See Section 3.5 for
details on setting alarm
limits.
OFF, except
SpO
2
low
which is 86
9b Pediatric Alarm Limits Set the default alarm limit
values for a Pediatric size
patient. See Section 3.5 for
details on setting alarm
limits.
OFF, except
SpO
2
low
which is 86
USER
CONFIGURATION
NUMBER FUNCTION DESCRIPTION
FACTORY
DEFAULT