Structure and Assembly/Disassembly 9-29
2. Pull out the plug shown in the figure and unscrew four M3X8 screws to remove touch
screen control board.
Touch screen
1. Remove 10.4 inch display module. Refer to 9.3.11.
2. Lift the edge of touch screen slightly to remove touch screen.
3. When assembling, fix touch screen into the slots as shown in the figure.
NOTE:
Please remove the protective film on the both sides before assembling touch screen and
make it clean . Make sure the touch screen is put into the “Llocating slot to prevent
damage
9.3.11.4 Minor Panel Six in One Encoder Module
1. Remove minor panel module. Refer to 9.3.11.
2. Pull out six covers of encoders.