5-86 Image Optimization
Tips:
In dual live mode, the screen displays the contrast image and tissue image
In freeze mode, there displays only one cine review progress bar as the contrast image
and tissue image are reviewed synchronously.
5.12.3.6 Mix Map
This function is to mix the contrast image with the tissue image, so that interested contrast regions
can be located.
Touch [Mix] to select different mixing mode, and select different maps by knobs under [Gray
Map]/[Tint Map]/[Mix Map] on the touch screen.
When dual live function is on, you can see the mixed effect on the contrast image.
When dual live function is off, you can see the mixed effect on the full screen image.
Select the map through the [Mix Map] item.
5.12.3.7 iTouch
On contrast status, you can also get a better image effect by using iTouch function.
Press <iTouch> on the control panel to turn on the function.
The symbol of iTouch will be displayed in the image parameter area in the upper right corner of the
screen once press <iTouch>.
Select different levels of iTouch effect through [iTouch] on the touch screen.
Long press the <iTouch> key to exit the function.
5.12.3.8 Image Saving
Live capture
In live mode, you can save the interested images by touching [Pro Capture] and [Retro
Capture].
Cine saving
In live mode, press the <Freeze> key on the control panel to enter cine review status.
5.12.4 Measurement, Comment and Body Mark
The system supports image measurement, comment and body mark functions. For the details,
please refer to the relevant sections.
5.12.5 Micro Flow Enhancement
MFE superimposes and processes multiple frames of contrast image during the cycle; it indicates
tiny vessel structures in detail by recording and imaging microbubbles.
MFE switch
During real-time scanning, touch [MFE] on the touch screen to start MFE imaging (MFE key is
highlighted in green).
Tips: MFE imaging is available only for real-time imaging or cine file of auto review mode. If a MFE
cine is reviewed manually, MFE effect cannot be displayed.
MFE Period
Rotate the knob under [MFE Period] on the touch screen to select different imaging period suitable
for current flow. Where, MAX is the maximum superimposing effect.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-80 and is the answer not in the manual?

Mindray DC-80 Specifications

General IconGeneral
Imaging ModesB, M, Color Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler, Tissue Doppler, 3D/4D Imaging
ProbesConvex, Linear, Phased Array, Volume
Advanced FeaturesAuto Optimization, Auto EF
ConnectivityDICOM 3.0, USB, HDMI, VGA, Ethernet
Transducer Ports4
ApplicationsAbdominal, Cardiac, Obstetrics/Gynecology, Vascular, Small Parts, Musculoskeletal, Urology, Pediatrics, Neonatal
Transducer TypesConvex, Linear, Phased Array, Volume
Power Supply100-240V AC, 50/60Hz
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 60°C
Humidity30% to 80%