Image Optimization 5-91
Exit
Press <B> or <Elasto> on the control panel to exit and enter the mode.
5.13.1.3 Pressure Hint Curve
The screen displays the pressure curve in real-time:
Where, the X-axis represents time and Y-axis represents pressure.
5.13.1.4 Parameter adjusting
ROI Adjustment
To adjust the width and position of ROI in Elasto imaging.
When the ROI box is solid line, roll the trackball to change its position.
When the ROI box is dotted line, roll the trackball to change the size.
Press <Set> to switch between the solid line and the dotted line status.
Smooth
To adjust the smooth feature of the Elasto image.
To adjust the image smooth, adjust through the [Smooth] item on the touch
screen.
Opacity
To adjust the opacity feature of the Elasto image.
Rotate the knob under [Opacity] item on the touch screen.
Invert
To invert the E color bar, thus to invert the color of benign and malignant tissue
can be inverted.
Click [Invert] on the touch screen.
Display Format
Adjust the display format of ultrasound image and the Elasto image.
Tap [H 1:1], [V 1:1], [Full] on the touch screen to adjust.
The system provides 3 types of display format:
H 1:1: right and left display (the real-time ultrasound image appears on the left,
and the elasto image appears on the right);
V 1:1: up down display (the elasto image appears above, and the real-time
ultrasound image appears below).
Full: the elasto image only displayed.
Map
To select different maps for observation.