Image Optimization 5-5 
5.2.3  B Mode Image Optimization 
Image Quality 
To switch between the fundamental frequency and harmonic frequency, and select 
the corresponding frequency type. The real-time value of frequency is displayed in 
the image parameter area in the upper right corner of the screen, and if harmonic 
frequency is used, ―F H‖ is displayed as harmonic frequency value. 
Press the knob under the [Image Quality] on the touch screen to be ―Harmonic 
On‖ to turn on the function. 
Rotate the knob under the [Image Quality] on the touch screen to select the 
different frequency values. 
The adjusting range of frequency values can be divided into 6 levels, penetration 
preferred, (HPen), general mode (HGen), resolution preferred (HRes), the mode 
between penetration preferred and general mode (HPen-Gen), the mode between 
general mode and high FR(HGen-FFR), high FR with resolution preferred(HRes-
FFR). 
Press the knob under the [Image Quality] on the touch screen to be ―Harmonic 
Off‖ to use the fundamental frequency. 
The adjusting range of frequency values can be divided into 3 levels, penetration 
preferred, (Pen), general mode (Gen), resolution preferred (Res). 
Rotate the knob under the [Image Quality] on the touch screen to select the 
different frequency values. 
The system provides an imaging mode using harmonics of echoes to optimize the 
image. Harmonic imaging enhances near field resolution and reduces low-
frequency and large amplitude noise, so as to improve small parts imaging. 
Please select the frequency according to the detection depth and current tissue 
features. 
 
Gain 
To adjust the gain of the whole receiving information in B mode. The real-time gain 
value is displayed in the image parameter area in the upper right corner of the 
screen. 
Rotate the <B> knob clockwise to increase the gain, and anticlockwise to 
decrease. 
The adjusting range is 0-100. 
Increasing the gain will brighten the image and you can see more received signals. 
However, noise may also be increased. 
 
Depth 
This function is used to adjust the sampling depth, the real-time value of which is 
displayed in the image parameter area in the upper right corner of the screen. 
Use the <Depth> deflector rod on the lower right part of the control panel to adjust 
the depth. 
The adjustable depth values vary depending upon the probe types.