47
Service Manual Combics
Display Cause Solution
Err101 - 104 Key is stuck Release key or contact your local Service Center
Key pressed whilst switching on
Err 320 Operating program memory faulty Contact your local Service Center
Err 335 Verified weighing platform not compatible Connect a compatible weighing platform
with the connected terminal
Err 340 Incorrect operating parameter (EEPROM) Turn scale off, then back on again.
If Err 340 remains displayed,
Contact your local Service Center
Err 341 Flash has lost data Turn scale off, then back on again.
If Err 340 remains displayed,
Contact your local Service Center
Err 343 Loss of data in the memory area for trans- Contact your local Service Center
action numbers in external alibi memory
Inf 01 Data output not compatible with Set output format correctly
output format
inf 02 Calibration/adjustment condition not met, Calibrate only when zero is displayed
e.g., the scale was not tared - the scale Unload scale, tare using the ) key
is loaded
inf 03 Adjustment could not be completed Allow to warm up again and repeat the
within a certain time adjustment process
Inf 06 Built-in calibration weight defective Contact your local Minebea Intec Service Center
Display Cause Solution
Operating Design | Error Description Display Combics | 1 | 2 | 3

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea Intec Combics Series and is the answer not in the manual?

Minebea Intec Combics Series Specifications

General IconGeneral
BrandMinebea Intec
ModelCombics Series
CategoryScales
LanguageEnglish

Related product manuals