6. Operating | 6.2 - Configuration and adjustment via the keypad | 111
[Cut-off frequency] CP 012 FCut
The smaller the cutoff frequency, the slower the measurement and the
more stable the measurement result.
Input: 0.1…80.00Hz
The available options depend on the measure time.
The parameter FCut can only be selected if the digital filter is switched on.
[Test mode] CP 013 tStMod
AbS (absolute) is used to calculate the test value when the test is called up.
Rel (relative) displays the deviation from the originally saved test value.
[Standstill time] CP 015 StStIM
The parameters [Standstill time] and [Standstill range] define the
standstill of the scale (stable balance).
Input: 0.01–2.00s; the standstill time cannot be smaller than the measure
time.
[Standstill range] CP 016 StSrnG
As long as the weight fluctuations remain within this range, the device is
determined to be stable.
Input: 0.01…10.00d
[Tare timeout] CP 017 TArtIM
Timeout for a tare/zeroset command that cannot be executed (e.g. due to
mechanical instability of the scale, incorrect filter setting, resolution too
high, standstill condition too strict).
Input: 0.1…25.0s.
[Zeroset range] CP 018 ZESrnG
Determine a ±range around the zero point determined by the dead load
during adjustment; within this range
– the respectively displayed gross weight can be set to zero by a corre
sponding external command and
– automatic zero tracking is switched on.
Input: 0.00…10000.00d
[Zerotrack indic. range] CP 019 ZEtrnG
Display area within which automatic zerotrack compensates for
deviations.
Input: 0.25…10000.00d