9. Fieldbus interface | 9.2 - Scale protocol | 159
9.2.5.3 Function number 2: state-controlled action bits (read)
Display of signal state
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2 87: Get
FixTare
86: Set
FixTare
85: Re
setPwf
84: Re
setTest
83:
SetTest
82: Re
setTare
81:
SetTare
80: Set
Zero
Byte 3 89: Re
setError
88: Set
Print
Note: See Chapter Function number 80–92: state-controlled action bits
(write).
9.2.5.4 Funktion number 3: state of edge-controlled action bits (read)
The state is always 0.
9.2.5.5 Function number 4: adjustment information, error byte (read)
Byte Description
0: EXPO One byte for the position of the decimal point; content in
decimal form: 0 to 255.
0 = 000000
1 = 00000.0
2 = 0000.00
3 = 000.000
4 = 00.0000
5 = 0.00000