Page #162 background image
9. Fieldbus interface | 9.2 - Scale protocol | 162
9.2.5.6 Function number 5: device type and software version (read)
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0 TYPE MSB
Byte 1 TYPE LSB
Byte 2 MAINVERSION
Byte 3 SUBVERSION
e.g.: PR 5310 Rel 1.23 = 53100123
hex
9.2.5.7 Function number 6: serial number of the weighing point (read)
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0 Serial number MSB
Byte 1 Serial number
Byte 2 Serial number
Byte 3 Serial number LSB
e.g.: 148388723 = 08D83B73
hex
9.2.5.8 Function number 7: For internal use only.