3.1 POWER OFF
The meter is powered off by first pressing the Select key so that the OFF selection is activated,
then pressing the Select key again and holding for four seconds while the counter counts
down.
If the Select key is released before the four second countdown and following beep, the meter
will not power off. Instead, it will return to the menu display OFF.
When powered off, the instrument does not completely shut down all the internal circuitries.
The RTC-circuit remains active maintaining the time signal. The energy is supplied with the
batteries. The time setting of the instrument RTC is lost in case the batteries are removed for
more than 3 hours. The time can be reset using the CSW-32 Configuration Software.
3.2 CUMULATIVE DOSE
The meter integrates the cumulative dose continuously when switched on. The cumulative
dose can be viewed by pressing the DOSE key in the Main display or by entering the menu
and navigating to DOSE.
Example of the Cumulative dose display
accessible via the DOSE key in the Main display.
The display features a percentage scale which
indicates the accumulated dose percentage with
respect to the dose alarm level setting.