MAINTENANCE AND DECONTAMINATION
41
5.2 BATTERY COVER AND SEAL CHANGE
For battery cover and seal change a replacement cover or seal and a Pozidriv #1
screwdriver are needed.
Instructions:
Unscrew the battery cover from the meter
Pull the rubber seal from the plastic cover.
Attach the new rubber seal to the plastic cover.
Make sure that the tips in the plastic are in the holes of the seal and the seal is well
aligned.
Screw the battery cover back to the meter.
5.3 CALIBRATION
The CSW-32 Configuration Software is needed for the calibration of RDS-32. Follow the
calibration procedure outlined in the software manual.
5.4 MECHANICAL DECONTAMINATION
The recommended mechanical method for decontamination is vacuum cleaning or blasting
with pressurized air.
5.5 CHEMICAL DECONTAMINATION
During the decontamination procedure following items are needed:
Cleaning solution with neutral pH (non-corrosive preferred)
Cotton pad, paper tissue, etc.
The RDS-32 can be decontaminated with any commercially available decontamination
solutions which are suitable for polycarbonate/polyamide materials (e.g. Sprint 200 Free PH
7). Acetone is not recommended. Begin with wiping and/or brushing. Immersion is only
recommended as a next step if necessary and after it has been confirmed that the IP67 sealing
of the meter is intact. Please see the service manual or ask the manufacturer for more advices
about immersing the RDS-32 in a decontamination solution.