ERROR CODES AND OTHER DISPLAY MESSAGES
38
4 ERROR CODES AND OTHER DISPLAY MESSAGES
When necessary the meter will give display messages regarding battery condition, current
alarms, warnings and error situations.
4.1 LOW BATTERY WARNING
An audible low battery warning signal combined with a blinking battery symbol indication
repeated every 6 minutes when the battery capacity is 10% < Capacity < 30%
Low battery warning:
_______ _ _______
(Long Short Long Bleep in five minutes
intervals). The dose rate measurement continues normally.
4.2 LOW BATTERY ALARM
The display shows "LOBAT" blinking and the audible alarm is activated when the battery
capacity is <10%.
Low battery alarm:
_______ _ _______
(Long Short Long Bleep in 20 second intervals).
NOTE: The dose rate measurement is prevented.
4.3 DOSE RATE / DOSE WARNING (BLINKING DISPLAY)
Dose rate warning: ------ (two beeps per second).
The dose rate display is blinking, and the continuous audible alarm is on. The alarm can be
silenced with right arrow key in the Main displa.
Dose alarm: - - - - - - (one beep per two seconds).
The dose rate display is on and the current dose warning level alternates in the display and
the continuous audible alarm is on. The alarm can be silenced with the right arrow key in the
Main display.