RDS-32 FUNCTIONALITY
15
Cancel the maximum recorded dose reset process by pressing the left arrow key.
3.4 CHIRP FUNCTION
The chirp function can be used to activate or deactivate the audible or visual chirp and to
select the chirp divisor. The audible/visual chirp is an indication of the number of pulses
received by the detector. The possible chirp divisors are: 1/1, 1/2, 1/5, 1/10, 1/20 and 1/50.
The audible and visual chirp can be disabled individually, but they share the common dividing
ratio when enabled. To see or change the chirp state navigate to CHIRP in the RDS-32 menu.
The menu shows AUD:ON or AUD:OFF depending on the current state of the audible chirp.
Change the setting by navigating into the Audio menu.
The RDS-32 records the measured maximum dose rate while switched on and saves that
measured value even when the meter is switched off/on. To follow the maximum dose rate
by task or mission, reset the record as indicated above. Alternatively, set a suitable
histogram data save period so that the maximum dose rate can be discerned from the
histogram data. The maximum dose rate is recorded with the same time interval as is set
in the histogram settings. See Chapter 3.9 Histogram for more details. Note that the CSW-
32 configuration software is required to access the histogram data.