RDS-32 FUNCTIONALITY
26
3.8.2 BLE
The device has readiness for Bluetooth communication. Contact Mirion for additional
information.
BLE communication can be configured in the BLE menu. The user can manually activate
BLE enumeration for 60 s. The user can also disable to BLE communication in this menu.
ON will activate BLE enumeration for 60 seconds
OFF will disable the BLE communication
If the BLE is activated the Main display will show the symbol .
3.9 HISTOGRAM
The default setting in the RDS-32 is that the histogram is OFF. The histogram collection can
be switched ON in the HISTO menu. By default the data collection is set at 10 second
intervals.
Note that the histogram data is only accessible via the CSW-32 Configuration Software. The
histogram memory is downloaded using the software and USB cable provided with the
software package. The stored data is in XML format.
The meter memory is able to store approximately 60000 data sets. The information stored is
the immediate value in the display, the average and maximum dose rate from the previous
storing, the location and time. When an external detector is connected to the instrument, the
histogram stores data from both internal and external detector.
The final number of stored data sets depends on the amount of diagnostics information that
is also stored into the same memory area.
It will take approximately 166 hours to fill the RDS-32 histogram memory when sampling at
10 second intervals. If the histogram memory becomes full the older data sets will be
overwritten by new data sets.