RDS-32 FUNCTIONALITY
20
3.5.2 DOSE ALARM
The dose alarm can have three different modes:
Single
Dual (different level is set for WARNING and ALARM)
OFF
To switch between the different alarm modes navigate to Dose Mode and set the desired
mode Dual, Single or OFF by pressing the Select key.
To set a new dose alarm level, navigate into Dose and set a value by pressing the up/down
arrow keys and then moving right (or left) between the numbers to complete your desired
setting. When the alarm level has been entered, save by pressing the Select key.
To set the desired alarm level, it may be necessary to switch between unit multipliers. To
switch between µSv, mSv or Sv, navigate to the unit (furthest right on the display) and change
the selected unit multiplier with the up/down arrow keys. When the desired alarm level is
selected, save by pressing the Select key.
To ignore the changes, press the left arrow key until the display returns to the previous menu
level. Also waiting 12 seconds time out will switch the display back to Main display without
changing the alarm level.
In Dual mode the user can set two levels which will trigger the device alarming functions: a
warning level and an alarm level.
To set the device in dual mode navigate into Dose Mode and select Dual from the menu. To
set a dose warning and alarm navigate into Dose. The first available setting will be Dose
Warning. Set a suitable level, lower than the intended alarm level, according to the process
described above. Save the desired level by pressing the Select key. The setting will move
automatically to Dose Alarm. Set a suitable dose alarm according to the process described
above.
If the dose warning is muted from the Main display, the warning audible, vibration and LED
indication will switch off. The display will still continuously show the warning symbol in the left
upper corner and the display will periodically indicate the dose warning level and the text
‘warning’.