RDS-32 FUNCTIONALITY
24
3.7 DIAGNOSTICS
The Diagnostics menu has the following features:
Manual activation of the meter selftest
Display the device firmware version
Display the battery capacity level
The device firmware can be updated using the CSW-32 Configuration Software. See the
software manual for more details.
To activate the meter selftest routine, navigate into the DIAG menu, the selftest is the active
selection and press the Select key to initiate.
The selftest routine includes checking the high voltage, the operating voltage and the battery
capacity. During this process the user may also check the operation of all the display pixels,
the visual alarm LED, the audible alarm and the vibration alarm.
When using NiMH batteries, set the correct battery type in the CSW-32 configuration
software. When using NiMH batteries, the battery capacity indication is less accurate
than for alkaline batteries.