RDS-32 FUNCTIONALITY
31
3.10.2 SCALER FUNCTION WITH OTHER EXTERNAL PROBES
With an external probe connected, the Scaler menu has the following features:
• Settings
• Gross Meas.
• Background
• Net Meas.
The Settings menu allows the user to change the probe unit and to set a measurement
integration time or number of counts.The preset time/count levels:
10 s, 1 min, 3 min, 6 min, 10 min, 30 min and 60 min or 100, 1000, 10000, 20000 counts.
The options available in the scaler menu will depend on probe type. For example, for the
GMP-25 probe there is a time setting available instead of time/count and the scaler menu will
show the current isotope in use.
The result of a Scaler measurement with
RDS-32 internal detector or with an
external dose rate probe will show the
dose R (P) and the time T.