RDS-32 FUNCTIONALITY
22
For contamination probes such as GMP-25 or ABP-150 the preset units of the external
alarm depend on the unit selected in the Probe menu: cps, cpm, dpm, Bq or Bq/cm2.
• CPS: OFF, 10, 15, 20, 60, 100, 60, 1000 or 6000 cps.
• CPM: OFF, 600, 900, 1200, 3600, 6000 cpm, 36 kcpm, 60 kcpm or 360 kcpm
• DPM: OFF, 1200, 1800, 2400, 7200 dpm, 12 kdpm, 72 kpdm, 120 kdpm or 720
kdpm
• Bq: 20, 30, 40, 120, 200, 1200, 2000 Bq or 12kbq
• Bq/cm2: list will be Bq values divided by detector area
For pulse rate probes like the GMP-11-3 only the cps and cpm units are available.
Set the probe alarm level by navigating into the EXT.AL menu and pressing the up/down
arrow keys until the desired alarm level is the active selection. Press the Select key to save
the desired alarm level.
Navigate into the Probe menu to change the probe measuring unit, if necessary. See
Chapter 3.11 External Probes for more details.