SPECIFICATIONS
43
RADIOLOGICAL CHARACTERISTICS:
RDS-32
RDS-32 iTx
RDS-32 WR
RDS-32 iTx WR
Operational quantity
Ambient dose equivalent
H*(10)
Ambient dose equivalent
H*(10)
Ambient dose equivalent
H*(10)
Ambient dose equivalent
H*(10)
Radiation detected: gamma and X-rays: Low
48 keV
48 keV
55 keV ǀ 65 keV**
55 keV ǀ 65 keV**
Radiation detected: gamma and X-rays: High***
1.8 MeV
1.8 MeV
1.8 MeV
1.8 MeV
Dose rate measurement range
0.05 μSv/h .. 100 mSv/h
0.005 mrem/h .. 10 rem/h
0.05 μSv/h .. 100 mSv/h
0.005 mrem/h .. 10 rem/h
0.05 μSv/h .. 10 Sv/h
0.005 mrem/h .. 1000 rem/h
0.05 μSv/h .. 10 Sv/h
0.005 mrem/h .. 1000 rem/h
IEC Dose rate measurement range: Low
0.3 μSv/h
0.03 mrem/h
0.3 μSv/h
0.03 mrem/h
0.3 μSv/h
0.03 mrem/h
0.3 μSv/h
0.03 mrem/h
IEC Dose rate measurement range: High
100 mSv/h
10 rem/h
100 mSv/h
10 rem/h
10 Sv/h
1000 rem/h
10 Sv/h
1000 rem/h
Dose measurement range: Low
0.1 μSv
0.01 mrem
0.1 μSv
0.01 mrem
0.1 μSv
0.01 mrem
0.1 μSv
0.01 mrem
Dose measurement range: High
10 Sv
1000 rem
10 Sv
1000 rem
10 Sv
1000 rem
10 Sv
1000 rem
Calibration accuracy at reference calibration direction and in the Cs-137 calibration field, temperature +20
°C (68°F). Compared to Finnish National Laboratory STUK.
± 5% @137Cs
± 5% @137Cs
± 5% @137Cs
± 5% @137Cs
Dose rate linearity
-15% - 22%
0.3 μSv/h...0.1 Sv/h
- 15%-22%
0.3 μSv/h...0.1 Sv/h
-15% -22%
0.3 μSv/h...10 Sv/h
- 15% - 22%
0.3 μSv/h...10 Sv/h
Variation of the response due to photon radiation energy and angle of incidence
(R E,A) 71% < RE,A <
167% (48 keV...1.8
MeV),
± 45°
(R E,A) 71% < RE,A <
167% (48 keV...1.8
MeV),
± 45°
(R E,A) 71% < RE,A < 167%
(55 keV ǀ 65 keV**...1.8
MeV), ± 45°
(R E,A) 71% < RE,A <
167% (55 keV ǀ 65
keV**...1.8 MeV), ± 45°
**for Si diode detector above 30 mSv/h dose rates
*** High energy response ratio for GM tube: 2.2 at 4.4 MeV and 2.6 at 6.7 MeV
High energy response ratio for Si diode: 1.2 at 4.4 MeV and 2.0 at 6.7 MeV