RADIOLOGICAL CHARACTERISTICS:
Ambient dose equivalent
H*(10)
Ambient dose equivalent
H*(10)
Ambient dose equivalent
H*(10)
Ambient dose equivalent
H*(10)
Radiation detected: gamma and X-rays: Low
Radiation detected: gamma and X-rays: High***
Dose rate measurement range
0.05 μSv/h .. 100 mSv/h
0.005 mrem/h .. 10 rem/h
0.05 μSv/h .. 100 mSv/h
0.005 mrem/h .. 10 rem/h
0.05 μSv/h .. 10 Sv/h
0.005 mrem/h .. 1000 rem/h
0.05 μSv/h .. 10 Sv/h
0.005 mrem/h .. 1000 rem/h
IEC Dose rate measurement range: Low
IEC Dose rate measurement range: High
Dose measurement range: Low
Dose measurement range: High
Calibration accuracy at reference calibration direction and in the Cs-137 calibration field, temperature +20
°C (68°F). Compared to Finnish National Laboratory STUK.
-15% - 22%
0.3 μSv/h...0.1 Sv/h
- 15%-22%
0.3 μSv/h...0.1 Sv/h
-15% -22%
0.3 μSv/h...10 Sv/h
- 15% - 22%
0.3 μSv/h...10 Sv/h
Variation of the response due to photon radiation energy and angle of incidence
(R E,A) 71% < RE,A <
167% (48 keV...1.8
MeV),
± 45°
(R E,A) 71% < RE,A <
167% (48 keV...1.8
MeV),
± 45°
(R E,A) 71% < RE,A < 167%
(55 keV ǀ 65 keV**...1.8
MeV), ± 45°
(R E,A) 71% < RE,A <
167% (55 keV ǀ 65
keV**...1.8 MeV), ± 45°
**for Si diode detector above 30 mSv/h dose rates
*** High energy response ratio for GM tube: 2.2 at 4.4 MeV and 2.6 at 6.7 MeV
High energy response ratio for Si diode: 1.2 at 4.4 MeV and 2.0 at 6.7 MeV