Chapter Two: Installation Product Location and Re
q
uirements
23
Product Location and Requirements
The Type 247 unit meets the following criteria:
POLLUTION DEGREE 2 in accordance with IEC 664
Transient overvoltages according to INSTALLATION CATEGORY II
Operating Environmental Requirements
Ambient Operating Temperature: 15
°
to 40
°
C (59
°
to 104
°
F)
Main supply voltage fluctuations must not exceed
±
10% of the nominal voltage
Ventilation requirements include sufficient air circulation
Connect the power cord into a properly grounded outlet
Safety Conditions
The 247 unit poses no safety risk under the following environmental conditions.
Altitude: up to 2000 m
Maximum relative humidity: 80% for temperatures up to 31
°
C, decreasing linearly to
50% at 40
°
C