Page12of86
thesystemforoptimalresults.Specifically,a5VDCwithacurrentcapacityof2A,regardlessofthe
powersupplyinput.
1.8. ListofPrecautions
Precautionsinthismanualregardthespecialcareexercisedbythepractitionerforthesafeandeffectiveuse
ofthecolposcope.
1.8.1. Donotallowthecolposcopetocomeincontactwiththepatient.
1.8.2. DonottouchtheopticallensorLEDlightexceptasdescribedinthemaintenancesectionofthis
manual.
1.8.3. Donotimmerseanypartofthecolposcopeincleaningsolutionsoranyfluid.
1.8.4. Donotusestrongdetergentsoralcoholunderanycircumstancesforopticscleaning.Thiswouldleadto
irrevocabledamagetotheopticallycoatedlenssurfaces.
1.8.5. Tolimitcontamination,userecommendeddisinfectantagentsbetweeneachexam.
1.8.6. Donotusecorrosivematerialsorstrongcleaningmaterialsotherthanthosespecificallyprescribedon
thecasingsorontheoptics.Suchmaterialsleadtopermanentdamageofthepaintedsurfaceofyour
colposcopeand/orscratchthelenscoatingsoftheoptics.Thisleadstoconsiderablelossofqualityof
theentirevisualsystem.
1.8.7. TheEVASystemwasbuilttoberesilient,butneedstobetakencareoflikeanyotherelectronic
device.Itshouldneverbestoredoroperatedinareaswhereitcouldgetwetorcouldbeexposedto
environmentalconditionssuchasextremetemperatureorhumidity,prolongeddirectsunlight,dust,
etc.
1.8.8. IftheLEDlightceasestowork,duetobatterydrainorLEDburnout,stopusingthecolposcopeand
rechargethebatteries.SometimesadischargedbatterywillmanifestviaaflickeringLED.Inthiscase,
rechargethebatterytofullcapacitybeforecontinuinguse.
1.8.9. Nomodificationofthisequipmentisrecommended.
1.9. ProductOverview
Beforeuseofyourcolposcope,checkfordamageontheoutsideofthetransportpackaging,theopticalunit,
mobiledevice,andanyaccessories.
Thefollowingchaptersprovidestepbystepinstructionsforthetesting,installation,operation,andsecurity
references.
1.10. Components
BeforeinitialsetupoftheEVASystem,checkthatyouhavereceivedallthefollowingcomponents:
Opticalunitbodyandlens
SamsungJ530mobiledevice*
Syncwire2Ports30W/6ADualPortMicroUSBCharger
Sparepartskitincluding:1chargingcable,2technicianscrewsandanAllenkey
Neckstrap
Lenscleaningcloth
Userguide
Ifanypartsaremissing,pleasenotifyMobileODT.
*PleaserefertotheSamsungJ530usermanualtofindspecificationsregardingtheWiFi,howthemobile
deviceshouldbeconfiguredandoperatedincludingthequalityofservice,operatingdistancesandranges,
securityrequirements,andhowtodealwithanyrisksandproblemsthatmayarise.