EVASystemUserManual
No.10LB038
Rev.04
Page31of86
range400700mm
Uselife Unitbatte
ars
8.1. Electromagneticcompatibilitymanufacturerdeclaration
8.1.1. TheEVASystemrequiresspecialprecautionsinregardtoelectromagneticcompatibility.
8.1.2. ItmustbeinstalledandpreparedforuseasdescribedintheOperationsectionofthisUserManual.
8.1.3. Theseparationdistancesrecommendedinthisparagraphmustthereforebecompliedwith.
8.1.4. TheEVASystemmustnotbeusednearorontopofanotherelectronicdevice.Ifthiscannotbeavoided,
itisnecessary–beforeclinicaluse–tochecktheequipmentforcorrectoperationundertheconditions
ofuse.
8.1.5. TheuseofaccessoriesotherthanthosespecifiedorsoldbyMobileODTasreplacementpartsmayhave
theconsequenceofincreasingtheemissionsordecreasingtheimmunityoftheunit.
8.2. Electromagneticemissions
8.2.1. TheEVASystemisintendedforuseintheelectromagneticenvironmentspecifiedinthefollowingtables
1,2,4and6below.
8.2.2. Theuserand/orinstalleroftheunitmustensurethatitisusedinsuchanenvironment.
Table1
Guidanceandmanufacturer'sdeclaration–electromagneticemissions–EVASystem
TheEVASystemisintendedforuseintheelectromagneticenvironmentspecifiedbelow;ThecustomerortheuseroftheEVASystem
shouldassurethatitisusedinsuchanenvironment.
Emissions
test
Compliance Electromagneticenvironmentguidance
RF
emissions
CISPR11
Group1
TheEVASystemusesRFenergyonlyforitsinternalfunction.Therefore,itsRFemissionsareverylow
andarenotlikelytocauseanyinterferenceinnearbyelectronicequipment.
RF
emissions
CISPR11
ClassB
TheEVASystemissuitableforuseinallestablishments,includingdomesticestablishmentsandthose
directlyconnectedtothepubliclowvoltagepowersupplynetworkthatsuppliesbuildingsused
fordomesticpurposes.
Guidanceandmanufacturer'sdeclaration–electromagneticimmunity–EVASystem
TheEVASystemisintendedforuseintheelectromagneticenvironmentspecifiedbelow;thecustomerortheuseroftheEVASystem
shouldassurethatitisusedinsuchanenvironment.
Immunity
test
IEC6060112
Testlevel
Compliancelevel Electromagneticenvironmentguidance
Doc ID: 4baaf484d7f5f01d2b8e5dc0fedc637256236abe