EVASystemUserManual
No.10LB038
Rev.04
Page32of86
Electrostati
cdischarge
(ESD)IEC
6100042
8kVcontact
15kVair
8kVcontact
15kVair
Floorsshouldbewood,concreteor
ceramictile.Iffloorsarecoveredwith
syntheticmaterial,therelativehumidity
shouldbeatleast30%.
Power
frequency
(50/60Hz)
magnetic
field
IEC61000
48
30A/m
30A/m
Powerfrequencymagneticfieldsshouldbeatlevelscharacteristicofatypical
publiclowvoltagepowersupplynetworkthatsuppliesbuildingsusedfor
domesticpurposes,commercialorhospitalandclinicalenvironments.
Guidanceandmanufacturer'sdeclaration–electromagneticimmunity–EVASystem
TheEVASystemisintendedforuseintheelectromagneticenvironmentspecifiedbelow;thecustomerortheuseroftheEVASystem
shouldensureitisusedinsuchanenvironment.
Immunitytest
IEC6060112
Testlevel
Compliance
level
Electromagneticenvironmentguidance
RadiatedRF
IEC6100043
10V/m
80MHzto2.5GHz
10V/m
PortableandmobileRFcommunicationsequipment
shouldbeusednoclosertoanypartofthe
EVASystem,includingcables,thenthe
recommendedseparationdistancecalculatedfrom
theequation,applicabletothefrequencyofthe
transmitter.
Recommendedseparationdistance
d=6/E√P80MHzto2.5GHz
wherePisthemaximumoutputpowerratingofthe
transmitterinwatts(W)accordingtothetransmitter
manufactureranddistherecommendedseparation
Distanceinmeters(m)andEistheimmunitytestlevelinV/m.
FieldstrengthsfromfixedRFtransmitters,as
determinedbyanelectromagneticsitesurvey,
a
shouldbelessthanthe
compliancelevelineachfrequency
range.
d
Interferencemayoccurinthevicinityofequipment
markedwiththefollowingsymbol:
Doc ID: 4baaf484d7f5f01d2b8e5dc0fedc637256236abe