56
EN
B CLASSIC / B FUTURA / S CLASSIC - TECHNICAL MANUAL
SERVICE MENU
SERVICE MENU – SN INPUT
SQ33
Function SN INPUT lets you store the unique serial number of each
individual autoclave.
The serial number is stored at the factory using commands + / - to
change:
- 2 gures that denote year of manufacture
- 2 progressive groups of gures
- and press OK to conrm
The two letters, unique to each model, are automatically assigned when
product code is entered (see REF INPUT).
SERVICE MENU – WATER FILLING TIMES
SQ34
Selecting the WATER FILLING TIMES menu displays the following
indicators:
- Max lling – maximum feed tank level
- Min lling – minimum feed tank level
- Outlet – supply to auto lling device
- Time – auto lling activation time
The automatic lling activation time is shown by the sliding bar found under
the Time indicator.
The two counters indicate the times expressed in seconds for:
Alarm – error indication during the rst automatic lling phase, from an
empty tank up to minimum level.
Filling – time duration of second tank lling phase, from minimum level to
full tank, without reaching maximum level.
The WATER FILLING TIMES menu shows the ALARM and FILLING times
stored at the factory for:
- external lling pump 150/150
- Pure 100 700/700
- Pure 500 1000/1000
- EV AUX 1000/1000