Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
98
ALARMS / ERRORS / WARNINGS / TROUBLESHOOTING
EN
CODE AND
MESSAGE ON
THE DISPLAY
DESCRIPTION
Possible causes /
Checks
Solution
ERROR
OUTPUT
(**)
A360
SPD PRESSURE
DOWNSTROKE
TIMEOUT
CONTACT TECH.
SERVICE
Chamber
depressurisation
not completed
within the timeout
Chamber
depressurisation not
completed within the
timeout
Clean the discharge filter (see
Maintenance Appendix in the user
manual).
2
A362
PD PRESSURE
DOWNSTROKE
TIMEOUT
CONTACT TECH.
SERVICE
Vacuum phase
error during
decreases in drying
cycle
Chamber
depressurisation not
completed within the
timeout
Clean the discharge filter (see
Maintenance Appendix in the user
manual).
(**)
1 = OK (warning) - Tap OK on pop-up window.
2 = OK + door unlock / RESET - Keep OK pressed for 3 seconds on the pop-up window and the autoclave will start
the procedure to open the door in full safety. When the RESET key appears in the left area of the LCD, hold it
pressed for 3 seconds.