Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
ALARMS / ERRORS / WARNINGS / TROUBLESHOOTING
CODE AND
MESSAGE ON THE
DISPLAY
E051
TEST REMINDER
RUN HELIX TEST
Reminder to run
Helix Test
Reminder to run Helix
Test cycle
Run Helix Test cycle as planned
E052
TEST REMINDER
RUN
VACUUM+HELIX
TEST
Reminder to run
Vacuum + Helix Test
Reminder to run
Vacuum + Helix Test
combined cycle
Run Vacuum + Helix Test
combined cycle as planned
E060
ETHERNET CONFIG.
ERROR
CHECK SETTINGS
Error during
connection to Lan
network due to
Ethernet
configuration
parameters
The autoclave cannot
connect to the Lan
network
Check that Lan network
configuration parameters are
correct. Check that Lan network
to which the connection is to be
performed is correctly working.
Connectivity module
problem
Replace the board on which
there is the connectivity module
E061
WI-FI CONFIG.
ERROR
CHECK SETTINGS
Error during
connection to Wi-Fi
network due to Wi-Fi
configuration
parameters
The autoclave cannot
connect to the Wi-Fi
network
Check that Wi-Fi network
configuration parameters are
correct. Check that the router
than manages the Wi-Fi network
is on and that the Wi-Fi network
to which the connection is to be
performed is correctly working.
Connectivity module
problem
Replace the board on which
there is the connectivity module
Update of the
connectivity module
started from USB or
remotely. The
message signals that
the update is in
progress
If the update from USB is not
successful, try to replace the
USB key