Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
EN
ALARMS / ERRORS / WARNINGS / TROUBLESHOOTING
83
CODE AND
MESSAGE ON
THE DISPLAY
DESCRIPTION
Possible causes /
Checks
Solution
ERROR
OUTPUT
(**)
A105
CONDUCTIVITY
SENSOR
PT5 OPEN CIRCUIT
CONTACT TECH.
SERVICE
PT5 probe open
circuit (conductivity
sensor)
PT5 wiring problem -
wiring check
Check connections:
- conductivity sensor --> PCB
1
PT5 open circuit
problem - probe check
Disconnect the conductivity sensor
from the PCB.
Detect continuity.
Replace the conductivity sensor
PCB problem
Replace the PCB
A111
CHAMBER PROBE
PT 1 SHORT
CIRCUIT CONTACT
TECH. SERVICE
PT1 probe short
circuit (chamber)
PT1 wiring problem -
wiring check.
Check the integrity of PT1 wiring,
paying special attention to the
connector on the PCB.
1
PT1 probe problem -
probe check.
Check PT1 resistive value
Replace PT1 probe - see technical
sheet ST05 - PT1 PROBE -
CHAMBER TEMPERATURE
READING
A112
GENERATOR
PROBE
PT2 SHORT-
CIRCUIT
CONTACT TECH.
SERVICE
PT2 probe short
circuit (steam
generator)
PT2 wiring problem
wiring check
Check the integrity of PT2 wiring,
paying special attention to the
connector on the PCB.
1
PT2 probe problem
probe check
Check PT2 resistive value.
Replace PT2 probe - see the
technical sheet ST01 - STEAM
GENERATOR.
A113
HEATING BAND
PROBE
PT3 OPEN CIRCUIT
CONTACT TECH.
SERVICE
PT3 probe short
circuit (tube bundle
heating element)
PT3 wiring problem -
wiring check
Check the integrity of PT3 wiring,
paying special attention to the
connector on the PCB.
1
PT3 probe problem -
probe check.
Check PT3 resistive value.
Replace PT3 probe
A115
CONDUCTIVITY
SENSOR
PT5 SHORT
CIRCUIT
CONTACT TECH.
SERVICE
PT5 resistance
thermometer in
short circuit
(conductivity
sensor)
PT5 wiring problem -
wiring check
Check the integrity of conductivity
sensor wiring, paying special
attention to the connector on the
PCB
1
PT5 probe problem -
probe check
Check resistive value.
Replace the conductivity sensor.
A116
PROCESS BOARD
ERROR
CONTACT TECH.
SERVICE
ADC error
ADC damaged
Replace the PCB
1