Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
EN
ALARMS / ERRORS / WARNINGS / TROUBLESHOOTING
85
CODE AND
MESSAGE ON
THE DISPLAY
DESCRIPTION
Possible causes /
Checks
Solution
ERROR
OUTPUT
(**)
A129
NFC MODULE
ERROR
CONTACT TECH.
SERVICE
Communication
error between
board L4 and NFC
module
Connection error with
NFC module
NFC module fault
Replace one PCB at a time until
the problem is solved
1
A130
RGB LED BAR
ERROR
CONTACT TECH.
SERVICE
Communication
error with RGB
LED bar on L4
Connection error with
RGB LED bar
RGB LED bar fault
Replace the LED bar and then the
PCB until the problem is solved
1
A131
SOLENOID VALVE
1 ERROR
CONTACT TECH.
SERVICE
Error from solenoid
valve driver on EV1
Solenoid valve 1
broken
Check the correct operation of the
solenoid valve from the Service
menu and replace it if necessary
2
PCB problem.
Replace the PCB
A132
SOLENOID VALVE
2 ERROR
CONTACT TECH.
SERVICE
Error from solenoid
valve driver on EV2
Solenoid valve 2
broken
Check the correct operation of the
solenoid valve from the Service
menu and replace it if necessary
2
PCB problem
Replace the PCB
A133
SOLENOID VALVE
3 ERROR
CONTACT TECH.
SERVICE
Error from solenoid
valve driver on EV3
Solenoid valve 3
broken
Check the correct operation of the
solenoid valve from the Service
menu and replace it if necessary
2
PCB problem
Replace the PCB
A134
SOLENOID VALVE
4 ERROR
CONTACT TECH.
SERVICE
Error from solenoid
valve driver on EV4
Solenoid valve 4
broken
Check the correct operation of the
solenoid valve from the Service
menu and replace it if necessary
2
PCB problem
Replace the PCB
A135
SOLENOID VALVE
5 ERROR
CONTACT TECH.
SERVICE
Error from solenoid
valve driver on EV5
Solenoid valve 5
broken
Check the correct operation of the
solenoid valve from the Service
menu and replace it if necessary
PCB problem
Replace the PCB
A136
SOLENOID VALVE
6 ERROR
CONTACT TECH.
SERVICE
Error from solenoid
valve driver on EV6
Solenoid valve 6
broken
Check the correct operation of the
solenoid valve from the Service
menu and replace it if necessary
2