EasyManua.ls Logo

Mocom SUPREME 17 - Page 90

Mocom SUPREME 17
128 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
88
ALARMS / ERRORS / WARNINGS / TROUBLESHOOTING
EN
CODE AND
MESSAGE ON
THE DISPLAY
DESCRIPTION
Possible causes /
Checks
Solution
ERROR
OUTPUT
(**)
A251
ATM1 UPSTROKE
TIMEOUT
CONTACT TECH.
SERVICE
The 1
st
pressure
pulse does not
achieve 0.00 bar
within timeout
Autoclave loading
problem - load check
Check quantity and arrangement
of the load inside the chamber.
Follow the instructions in the User
Manual.
2
Steam generator water
injection pump problem
- device check.
Steam generator
clogging problem -
device check.
Check:
- filling tank filter and water
injection pump inlet filter
- correct activation of water pump
and opening of EV6
For the check, it can be useful
to activate the water pump and
EV6 by means of the
SERVICE/COMPONENT
TEST/OTHER COMPONENTS
menu, which allows to detect air
passage through the generator
to the chamber
- presence of any leakage on the
water pump and generator
connecting pipes
Replace vibration pump
Perform steam generator
maintenance - see sheet ST01
Generator heating
problem - device check
Check the safety thermostat and
reset it if necessary.
Check steam generator heating
element.
See ST01 - steam generator
sheet.

Table of Contents

Related product manuals