Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
ALARMS / ERRORS / WARNINGS / TROUBLESHOOTING
CODE AND
MESSAGE ON
THE DISPLAY
A258
3PP UPSTROKE
TIMEOUT
CONTACT TECH.
SERVICE
The 3
rd
pressure
pulse does not
achieve 1.15 bar
(121°C) or 2.15 bar
(134°C) within
timeout
Autoclave loading
problem - load check
Check quantity and arrangement
of the load inside the chamber.
Follow the instructions in the User
Manual.
Steam generator water
injection pump problem
- device check.
Steam generator
clogging problem -
device check
Check:
- filling tank filter and water
injection pump inlet filter
- correct activation of water pump
and opening of EV6
For the check, it can be useful
to activate the water pump and
EV6 by means of the
SERVICE/COMPONENT
TEST/OTHER COMPONENTS
menu, which allows to detect air
passage through the generator
to the chamber.
- presence of any leakage on the
water pump and generator
connecting pipes
Replace vibration pump
Perform steam generator
maintenance - see sheet ST01
Steam generator
heating problem -
device check
Check the safety thermostat and
reset it if necessary.
Check steam generator heating
element.
See ST01 - steam generator
sheet.
Manifold solenoid valve
sealing problem -
component check
Check:
- correct power supply and
sealing of EV1
- EV3 sealing
MANIFOLD - see sheet ST02
Steam leakage from
door seal
Thoroughly clean the door seal
and the steel door leaf.
Replace the door seal.
Door sealing problem -
closing system check
Check:
for any steam leakage from the
door
- that the door seal is intact and
correctly positioned
- the correct operation of the door
closing and locking system
For the check, it can be useful to
activate the closing system by
means of the SERVICE /
COMPONENT TEST / DOOR
menu, detecting the current value
required by the tightening
servomotor.